Low SNR for Head and Body SPT due to setting of rrfCalibFlag (aka Calibration Type) in MRconfig.cfg incorrectly set to z…
Low SNR for Head and Body SPT due to setting of rrfCalibFlag (aka Calibration Type) in MRconfig.cfg incorrectly set to zero.
Symptom
Low SNR for Head and Body SPT. Values were approximately half normal with noise and signal both contributing. When looking through error log noticed error 2248226 See Long Description Low SNR for Head and Body SPT. Values were approximately half normal with noise and signal both contributing. When looking through error log files noticed error 2248226, see below. Error: 2248226 File: APS:RrfCalibrator::calibrate Based on the calibration flag, calibration values will be loaded from defaults
Error message
2248226 Based on the calibration flag, calibration values will be loaded from defaults.
Applies to
- Product
- 3T Eclipse For E2 software
- Software
- E2 3T Eclipse
- Area
- MR
Full cause & step-by-step resolution