Pular para o conteúdo principal
GE Healthcare · CTDiscoveryVCT200109106

SmartPrep not starting, red boxes showing in Prescription screen.

Symptom

SmartPrep not starting, red boxes showing in Prescription screen.

Error message

Error: 200109106 The current prescription requires XX number of scan keys for scan files. See long description for details Tue Aug 14 16:28:25 2007 Error: 200109106 Host: cccgvct1 Process: scanRx File: SIRx.cxx Line: 6802 Function: Data Acquisition : OC Processing The current prescription requires XX number of scan keys for scan files. The system currently has 0 scan keys available. You must change the prescription, wait for the current recons to complete, or release scan files to continue.

Error 200109106 You must change the prescription, wait for the current recons to complete, or release scan files to continue. See long description Wed Jul 22 08:39:53 2009 Error: 200109106 Host: VCTHD01 Process: scanRx File: SIRx.cxx Line: 8632 Function: Data Acquisition : OC Processing The current prescription requires 42 scan keys for scan files. The system currently has 28 scan keys available. You must change the prescription, wait for the current recons to complete, or release scan files to continue.

Applies to

Models
Discovery, VCT
Product
Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
08MW39.4 for Discovery CT750 HD 08MW39.4 for Discovery CT750 HD 11/17/2008 Inital M3 release. 08MW39.4 for Discovery CT750 HD 11/17/2008 Inital M3 release. - Discovery CT750 HD Collector 5323588 Rev 3 includes OS and Application CD's - Operating System Linux 5.3.8 Software CD 5323584 Rev 1 - Application CT750 HD DVD 5323585 Rev 3 ======================================================= - Class C CT750 5233586 Rev 1 - Class M CT750 5323587 Rev 1 An On-site / Remote identification aid . (Example: check_config) Result from check_config or swhwinfo: 08MW39.4.V40_PS_HD64_G_GTM 08MW39.4.V40_PS_HD64_G_GTL
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.