Will Not Scan with Gradient Errors due to Open Y-Axis BRM Winding.

Symptom

Will Not Scan with Gradient Errors.

Error message

2224583 Auto Prescan Failed, Possible Data Acquisition Failure. File: APS:aps_cf Line: 15 Auto Prescan failed, possible data acquisition failure.

2224666 Auto Prescan failed, possible data acquisition error. Try again or use Manual Prescan. Wed Dec 1 09:09:30 2004 Error: 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1322 Auto Prescan failed, possible data acquisition error. Try again or use Manual Prescan.

2224666 Auto Prescan failed, Gradient amplifiers are not ready

2241520 The GIP reported a Y Axis Current Distortion Fault. Wed Jul 15 09:01:19 2009 2241520 lx-mr NSP HPC_PORT_ID:GAP/GIP/GP 2232 The GIP reported a Y Axis Current Distortion Fault. Another fault reported in the log could be the cause of this fault. Possible causes: GRAM Tuning, miniGRAM, components of the Y Axis GPM, GIP, or cables.->

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion BRM 2144093-3 See long description for SMILE on returning BRM's. See comments in resolution concept for second opinion requirements before ordering this part SMILe 2/7/2006 Subject: Failed and unused BRM coils must be returned promptly for evaluation and repair. Effectivity: All Signa 1.5T Horizon, LX, EXCITE and HD MR systems. Problem: Unused BRM's are remaining on site too long which severely limits the availability of FRU BRM coils. The repair facility is able to perform quick evaluations/repair of coils but BRM's must be returned promptly. Resolution: BRM coils should be ordered with no more than a 3-day lead time . Every effort is being made to always have FRU BRM's available but your help is needed. Within 10 days of shipment of the FRU BRM either it or the failed unit must be returned. Any exceptions to this time limit must be cleared through Ramamohan Pai, Global Modality Leader-MR & XR. Email : [email protected] Tel: 1-262-544-3240, Dial Com : 320-3240
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.