LightSpeed 5.X (80/100kW)
Symptom
"Scanner Hardware Stopped Scan" DURING SMARTVIEW prep WITH Interventional Hand Held Control (IHHC) failed to enable Scanner Hardware Stopped Scan DURING SMARTVIEW prep WITH Interventional Hand Held Control (IHHC) failed to enable See the log description for the full error code Mon Sep 19 10:35:34 2005 Error: 182700 Suite: CT01 Host: ETC Hand Held Control Process: Diag Exec File hc_e xec.c 1.38 No Method Line: 595 Function: Diagnostics OPERATOR INPUT/OUTPUT : Tableside Operator Interface Test Name: Unique ID: 58982 Iteration: 1 Exception Level: Hard Time: 18945 Error Type: Firmware Interventional Hand Held Control (IHHC) failed to enable. Verify IHHC is connected to scanner. If IHHC is not connected to scanner, connect IHHC to scanner. If IHHC is connected to scanner, disconnect IHHC from scanner and reconnect IHHC to scanner. Thu Aug 14 12:26:27 2008 Error: 182701 Host: Process: gr:RECEIVE:firmware File: hc_wdog.c 1.7 2/16/99 Line: 255 Function: No System Function Reported Scan: 25796/3/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 12:26:26:920 Log Series: 418 Interventional Hand Held Control (IHHC) Watchdog Timeout. The IHHC has been disabled. HW/SW Error: Probable hardware, possible software, problem
Error message
Error: 260111604 Tx retries used. See long description for a full sample error. Wed Jul 2 11:44:36 2008 Error: 260111604 Host: tgp Process: tgp subsys File: hc tx.c Line: 143 Wed Jul 2 11:44:36 2008 Host : Table Gantry Processor Ermes # : 260111604 Exception Class : Abort Severity : Primary File : hc tx.c Line# : 143 Function : No System Function Reported Scan Type : Axial Scan: 4239/3/1 Tx retries used.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2199947 CT SONATA HAND HELD CONTROLLER CT SONATA HAND HELD CONTROLLER Part number 2199947 CT SONATA HAND HELD CONTROLLER Part number 2199947
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.