Three 3-Plane Localizers Produce Only Three 3 Images Per Plane and Scans Prescribed From Them Abort With SCT Interface E…

Three 3-Plane Localizers Produce Only Three 3 Images Per Plane and Scans Prescribed From Them Abort With SCT Interface Errors or “Infiniband backlog too large…scan stopped” Errors due to Obsolete Infi…

Symptom

Three 3-Plane Localizers Produce Only Three 3 Images Per Plane and Scans Prescribed From Them Abort With SCT Interface Errors See Long Description FMI 60753 (14.0 M5A software upgrade) installed Thursday evening, January 15, 2009. Upon the customer beginning their scanning day on Friday, January 16 they noticed the following problem: The 3-Plane Loc's produced only one image per plane, and when they tried to prescribe off these images and scan the first axial scan, the scans would always abort with the SCT Interface Errors as shown below. iTrak # 13236909. Fri Jan 16 08:25:35 2009 2248370 mcoamr1 NSP SCP:SctPxProxy 833 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Gradient Echo Coil: QUADANKLE Scan: 5207/1 SCT Interface Error. No source was detected for a SCT command. Fri Jan 16 08:25:35 2009 2248371 mcoamr1 NSP SCP:SctPxProxy 854 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Gradient Echo Coil: QUADANKLE Scan: 5207/1 SCT Interface Error. No timestamp was detected for a SCT command. Fri Jan 16 08:25:44 2009 2218537 mcoamr1 scn MsaScanMgr.m 3038 Scan stopped, Scan Stopped due to error : SCT Interface Error. No source was detected for a SCT command..

Error message

2218537 Scan stopped, Scan Stopped due to error SCT Interface Error. No source was detected for a SCT command. See Long Description 2218537 mcoamr1 scn MsaScanMgr.m 3038 Scan stopped, Scan Stopped due to error : SCT Interface Error. No source was detected for a SCT command..

Infiniband backlog too large…scan stopped See Long Description “ Infiniband backlog too large…scan stopped " Message is seen in the APS window.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR