GE Healthcare · MRSigna222853722185372225536
Cannot Scan Due to TR and Dynamic Disable Faults.
Symptom
Cannot Scan With intermittent TR and Dynamic Disable Faults. The customer when scanning is getting intermittent TR and Dynamic Disable faults. It happens with all coils and cannot be pinned down to a specific PSD.
Error message
2228537 Scan Stopped, A T/R or Dynamic Disable Fault Occurred Tue Apr 27 16:25:26 2004 Error: 2218537 Host: EMR1 Process: scn File: MsaScanMgr.m Line: 2358 Scan stopped, A T/R or dynamic disable fault occurred. 2225536 Dynamic disable and T/R driver board error. A spurious dynamic disable fault occurred. See Long Description Error: 2225536 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID PMH.m 1.29 Method: spi_tr_mon Line: 6220 Dynamic disable and T/R driver board error. A spurious dynamic disable fault occurred. It cleared itself before the specific problem could be read. Possible causes: noise on the dynamic disable board or a bad board.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.