GE Healthcare · MRSigna224530916278472253851
Gradient Diagnostic Failure due to STIF or SRF/TRF I/O Board Failure.
Symptom
Gradient Diagnostic Failure
Error message
2245309 Gradient Diagnostic test has been aborted because GP detected a bad clock input 2245309 WCLAGEMR2MR diagCoord grad_lib.c 924 Diagnostic: Grad Frame/Clk Iteration: 25 of infinite Action: 1 Task ID: 0x3fc4380 Time: 1627847 Object: SCP Test Task Function: Gradient: N/A Gradient Diagnostic test has been aborted because GP detected a bad clock input.-> Extended error log: "Check placement of Clock and X,Y,Z data fibers between the STIF and GP (look for swapped fibers between clock and x/y/z data fibers)" 2245309 WCLAGEMR2MR diagCoord grad_lib.c 924 Diagnostic: Grad Frame/Clk Iteration: 25 of infinite Action: 1 Task ID: 0x3fc4380 Time: 1627847 Object: SCP Test Task Function: Gradient: N/A Gradient Diagnostic test has been aborted because GP detected a bad clock input.-> Extended error log: "Check placement of Clock and X,Y,Z data fibers between the STIF and GP (look for swapped fibers between clock and x/y/z data fibers)" 2253851 GP Framing Error/Clock Stop Test Failure. The GP Failed to Detect a Framing Error During each pass of this test.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.