Symptom

No flashing eights on gantry display and Scanning hardware reset failed

Error message

Error: 182011 Final report watchdog time-out for notify group 10001 See the long description for the full error code Mon Aug 1 06:01:24 2005 Error: 182011 Host: ETC Process: Local Exc Handler File: exl notify.c Line:1032 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 06:01:23:990 Log Series: 33 Final report watchdog time-out for notify group 10001.

Error: 182004 LAN Watchdog TIMEOUT See LD for the full error code Mon Feb 14 10:28:30 2005 Error: 182004 Suite: SJ01 Host: ETC Process: Local Exc Handler File: exl_lan_in_mail.c Method: No Method Line: 175 Function: No System Function Reported Scan: 50238/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 10:28:30:550 Log Series: 31 LAN Watchdog TIMEOUT. HW/SW Error: Did not receive LAN Watchdog message from STC. (Current Time:431109 Last Wdog LAN Probe Time:430857 (in Ticks))

Error: 244608 some controllers(ETC, STC, OBC) do not response to ping. See long description for full error Wed Mar 25 06:59:25 2009 Error: 244608 Host: ggh-qxi Process: fwmgr File: HwManager.c Line: 1536 HARD RESET through serial connection failed. Some controllers(ETC, STC, OBC) do not response to 

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.