Pular para o conteúdo principal
GE Healthcare · CT186330

CT/e, CT Prospeed AI/FI Irix or Linux

Symptom

Axial faults with servo amp overcurrent. Overcurrent LED lights up on power up on the servo amplifier

Error message

Error:186330 Code:13-1002-06 Error detected by:TGP(GP) No feedback from Axial motor encoder was detected after Gantry rotation request. See Long Description Mon Apr 21 22:22:59 2008 Error: 186330 LINUX tgp_out tgp_errors.c 338 Gantry Error. Code : 13-1002-06 Error detected by : TGP(GP) No feedback from Axial motor encoder was detected after Gantry rotation request.

Applies to

Product
CT/e, CT Prospeed AI/FI Irix or Linux See long description for sale information Product Name: CT/e Single CT Scanner: Irix Platform A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) 3.01 / 3.02 / 3.03 Linux CTe Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found in the attachment Always use the most appropriate documentation for your system.
Component
2241899 Axial Servo Amp - used on YMS Hispeed with Performix Tube and CTe See long description for details Part #: 2241899 Part Type: GP CONSUMABLES Part Weight: 7.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9237SB SV AMP FRU: Yes Status: Orderable Comments: FLS - SERVO AMP U/O LXI,FXI,DXI PER PCN CT-979NP 2241899 BACKWARD COMPATIBLE TO 2189124. MS08992 DO NOT DISCOUNT - PER SHAKUN GELS - REASON A CURRENT MOTOR DRIVER (2189124) DOESN'T HAVE ENOUGH MARGIN FOR A REGENERATIVE BREAKING. PROCEDURE,AXIAL SERVO AMP LT1
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues