Symptom
Unable to turn on alignment lights. The gantry would not rotate to position to turn on the alignment lights Unable to turn on alignment lights. The gantry would not rotate to position to turn on the alignment lights.
Error message
Error:186330 Code:13-1002-06 Error detected by:TGP(GP) No feedback from Axial motor encoder was detected after Gantry rotation request. See Long Description Mon Apr 21 22:22:59 2008 Error: 186330 LINUX tgp_out tgp_errors.c 338 Gantry Error. Code : 13-1002-06 Error detected by : TGP(GP) No feedback from Axial motor encoder was detected after Gantry rotation request.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed Dual, ProSpeed FII Series HiSpeed Dual, ProSpeed FII Series Product Name: Hispeed Dual ( Linux ) CT/Be2 A list of types. (a Product/hardware table) This console has an HP8000, HP8200, or HP8400 Computer (not GRE console) An On-site / Remote identification aid . (Example: check_config) HP Host computer A Table of options (included, optional, new introductions, …) HiSpeed Dual for China and Global B71402CA has the following as standard Options 1.5 sec fast scan, Power 200ma, Thichness 1mm, Raw data 5000, Image 100000 A list of Software that has run on this product (Product / Software Compatibility Table) With HP8000/8200 : • OS Version 4.5.15144980•Software Version 2.235165193 (except China)•Software Version 2.235172510 (for China only)•Patch 2.245179238•Patch 2.255197828•Patch 2.275267772 With HP8400 : •OS Version 6.2.95265475•Software Version 3.105270457•Patch 3.305267336•DVD Interchange 6.15270187•Patch 3.40 (See Note 8)5316222•DVD Interchange 7.05316345N/ Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links: Drawing Print 2320306-2CCT : http://gemspdm.med.ge.com/ematrix/common/eLibDocument.jsp?type=type_DrawingPrint&name=2320306-2CCT&rev=9 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2241899 Axial Servo Amp - used on YMS Hispeed with Performix Tube and CTe See long description for details Part #: 2241899 Part Type: GP CONSUMABLES Part Weight: 7.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9237SB SV AMP FRU: Yes Status: Orderable Comments: FLS - SERVO AMP U/O LXI,FXI,DXI PER PCN CT-979NP 2241899 BACKWARD COMPATIBLE TO 2189124. MS08992 DO NOT DISCOUNT - PER SHAKUN GELS - REASON A CURRENT MOTOR DRIVER (2189124) DOESN'T HAVE ENOUGH MARGIN FOR A REGENERATIVE BREAKING. PROCEDURE,AXIAL SERVO AMP LT1
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.