GE Healthcare · X-ray R&FDefinium1600140

has no image after exposure intermittent

Symptom

has no image after exposure intermittent Ethernet communication problem between Magic PC and Detector such as Bad Ethernet cable or loose connection may happen sometime and result in issues with multiple instances of image buffer loss from detector. We need a method to check the state of the Ethernet communication

Error message

Error coad 1600140 Aborting image processing for the specified frame

Applies to

Models
Definium
Product
Definium 6000
Component
Replace the Ethernet cable PN# 5148115 FE name: Feixiaoyang MB:13511061596
Area
X-ray R&F

Full cause & step-by-step resolution