GE Healthcare · X-ray R&FDefinium1600140
has no image after exposure intermittent
Symptom
has no image after exposure intermittent Ethernet communication problem between Magic PC and Detector such as Bad Ethernet cable or loose connection may happen sometime and result in issues with multiple instances of image buffer loss from detector. We need a method to check the state of the Ethernet communication
Error message
Error coad 1600140 Aborting image processing for the specified frame
Applies to
- Models
- Definium
- Product
- Definium 6000
- Component
- Replace the Ethernet cable PN# 5148115 FE name: Feixiaoyang MB:13511061596
- Area
- X-ray R&F