Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2001090122001070232001091002100004413863552200502005

LightSpeed 4.X - GRE

Symptom

Unable to scan axial or helical; system cannot allocate image space as one chunk. Unable to scan axial or helical; system cannot allocate image space as one chunk

Error message

Error: 2106 Image Storage Full See Long Description for complete error Wed Aug 30 16:07:53 2006 Error: 2106 Host: IRIS Process: imserver File: imsprim.c Line: 742 Image Storage Full

Error: 200109012 Image Allocation Failed. Cannot build scan request. See Long Description for complete error Wed Mar 2 18:39:33 2005 Error: 200109012 Host: ct01 Process: scanRx File: UIRxToScanner.cxx Line: 1307 Function: Data Acquisition : OC Processing Image Allocation Failed. Cannot build scan request.

Error: 200107023 A Failure had been detected while calling UIR x_ToScanRequest function or srx_sm_generate_action function See Long Description for error variations Tue Oct 19 09:22:20 2004 Error: 200107023 Host: dgct Process: scanRx File: srx_sma_manage_scan.c Line: 688 Function: Data Acquisition : OC Processing srx_sma_prepare_send_request: A Failure had been detected while calling UIRx_ToScanRequest function. Nov 22 08:55:38 2006 Error: 200107023 Host: ct01 Process: scanRx File: srx_sm_event_notify.c Line: 148 Function: Data Acquisition : OC Processing srx_sm_event_notify: A Failure had been detected while calling srx_sm_generate_action function. Tue Aug 14 16:28:25 2007 Error: 200107023 Host: ccc

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues