Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT26011836733328118796926011827550108

LightSpeed 5.X (80/100kW)

Symptom

Scanner Hardware Stopped Scan with Tube casing temperature has exceeded 92 degree - document in RFS - critical issue Scanner Hardware Stopped Scan Tube casing temperature has exceeded 92 degree C.

Error message

Error: 260118367 - An unexpected JEDI state change occured - JEDI not ready See long description for full error Thu Jun 12 07:08:56 2008 Error: 260118367 Host: orp Process: orp_subsys File: gen_exec.c Line: 863 Thu Jun 12 07:08:55 2008 Host : On-Board Rotating Processor Ermes # : 260118367 Exception Class : Secondary Severity : Pri/Soft File : gen_exec.c Line# : 863 Function : No System Function Reported Scan Type : Scout Scan: 33328/1/1 An unexpected JEDI state change occured . JEDI not ready -- Tcks: 1187969 Line: 7420

Error: 260118275 (110-0903) Tube casing temperature has exceeded 92 degree C See LD for more details Sun Nov 6 12:44:33 2005 Error: 260118275 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Sun Nov 6 12:44:32 2005 Host : Jedi Ermes # : 260118275 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50108/1/3 (110-0903)Tube casing temperature has exceeded 92 degree C.^C Z6,Z7= 0h,0h

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2320973 H4 POWER I/F see long description for full description 2320973 H4 POWER I/F Part #: 2320973 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H4 POWER I/F FRU: Yes Status: Orderable Parts Offered as: Current : 2320973 Harvested : Not Available Exchange : Not Available Comments: FLS - H4 POWER I/F. ADDED PER THE HPOWER RSPL. EXPECTED M3 FOR THIS PROGRAM: FW19 2003. MS20544 ON 2/25/03. USED ON HP60 5151365 Hercules Heat Exchanger 48 Volt Power Supply Assembly with Cover See LD for more information Part #: 5151365 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Hercules Heat Exchanger Power Supply Assembly with Cover, part numbers 2306440 and 2298317 FRU: Yes Status: Orderable Parts Offered as: Current: 5151365 Harvested: Not Available Exchange: Not Available Comments: Added per ECR 2019074, this part is Power supply for VCT and HPower Heat Exchangers, ms13452 051120 POWER SUPPLY MANUFACTURED BY TOTAL POWER INTERNATIONAL INC. 418 BRIDGE ST. LOWELL, MA. 01850 350 WATT SINGLE OUTPUT SWITCHING POWER SUPPLY, UNIVERSAL INPUT, U- BRACKET 48 VOLT, 7.3 AMP PART NUMBER TPS350-14
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.