Hardware reset fails with Laser indicator flashing on the Gantry Display board

Symptom

Hardware reset fails with Laser indicator flashing on the Gantry Display board. Hardware reset fails with Laser indicator flashing on the Gantry Display board

Error message

Error: 244634 Subsystem State; DCB:down, CCB:down, TCB:down, JEDI:down, TGP/ORP:down, DEVICES:up. GOC6 version All Down except Devices See Long Description Mon Dec 8 14:32:41 2008 Error: 244634 Host: CTUH07 Process: fwmgr File: fwmgr_utils.c Line: 677 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:down CCB:down TCB:down JEDI:down TGP/ORP:down DEVICES:up . The error above is from a GOC6, devices up is new and this stands for new feature like intergrated injector which has not been release at this time ( 12/23/2008)

Applies to

Models
Discovery, VCT
Product
Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5264840 (Replaces 5180460) GFI Ground Fault Interupter Switching Supply Immune 30A 30mA_15mS Part #: 5264840 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Cutler-Hammer GFI Ground Fault Interupter Switching Supply Immune 30A 30mA_15mS FRU: Not Available Status: Orderable ====================== Parts Offered as: Current: 5264840 Harvested: Not Available Exchange: Not Available Comments: Per ECR 2067866, 212008898, 7 April 2011 OLD Part #: 5180460 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GFI Ground Fault Interupter Switching Supply Immune 30A 30mA_15mS FRU: Not Available Status: Orderable Comments: Per ECR 2037509, 212008898, 1/16/07 5180460 replaces 5126460
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.