GE Healthcare · MRSigna223849322386452238415
CWA ID13861. Clarify System Will Not Scan Due To Gradient Errors.
Symptom
Will Not Scan with Gradient Errors.
Error message
2238493 The MIF is reporting a long overload on the Z axis LOW 8645 Power Module. Thu Aug 14 09:54:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2238493 File: PMH.m 1.29 Method: gap Line: 1034 The MIF is reporting a long overload on the Z axis LOW 8645 Power Module. The LED on the 8645 for this fault should be lit. If the LED is not lit, check power and cable connections. To clear the fault, power cycle the Z axis LOW 8645 Power Module. 2238645 The Z axis GRAM is reporting an under voltage fault. Thu Aug 14 09:54:12 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2238645 File: PMH.m 1.29 Method: gap Line: 1045 The Z axis GRAM is reporting an under voltage fault. Most likely cause is PSD prescription with demanding parameters or GRAM Hardware Fault. 2238415 Z Axis Gradient Overload Occurred. The Z axis LOW 8645 Power Module has reported an overload. Aug 14 09:53:56 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2238415 File: PMH.m 1.29 Method: gap Line: 1061 Z Axis Gradient Overload Occurred. The Z axis LOW 8645 Power Module has reported an overload. An overload is the result of the power module output voltage not following the input command.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.