Unable to Scan with PDU Error due to the Compact PDU's Internal 24 Volt Power Supply Circuit Breaker Had Tripped.

Symptom

Unable to Scan with PDU Error

Error message

2241518 The GIP reported a GPM Power Supply Fault. 2241518 The GIP reported a GPM Power Supply Fault.File: HPC_PORT_ID:GAP/GIP/GP Line: 2359 The GIP reported a GPM Power Supply Fault. Check the GPM or the GPM Power Supply.->

2225697 MDS link failure. The IPG failed to receive an MDS transmission echo on a second try. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. MDS timer: 0x0012 Transmitted byte: 0x11 2225697 MDS link failure. The IPG failed to receive an MDS transmission echo on a second try. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. MDS timer: 0x0012 Transmitted byte: 0x11

2241508 The GIP is reporting ? Axis GPM Power Off Wed Jun 30 07:42:27 2004 Error: 2241508 Host: SJHMR Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 303 The GIP is reporting X Axis GPM Power Off. If all 3 axes are reporting this fault, check the GPM Power Supply circuit breakers and cables. Check GPM fuses and the cable between the GIP and the GPM.

2225336 The SPI received a packet indicating that SRI has reset. See Long Description Sun May 24 15:28:19 2009 2225336 MR4 NSP HPC_PORT_ID:spi_sri_recov

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR