GE Healthcare · Nuclear Medicine75928
No gantry or table motions with the Remote Control Unit (RCU) after resetting Ventri system
Symptom
No gantry or table motions with the Remote Control Unit (RCU) after resetting Ventri system Getting a message on the acquisition screen that said the RCU was not connected. No gantry or table motions with the Remote Control Unit (RCU) after resetting Ventri system Getting a message on the acquisition screen that said the RCU was not connected.
Error message
CRC Error: RCU reception support in inconsistent state ERROR=Error in GCU .../gcu/hwcomponents/CardiacSignalHandler.cpp(307)::75928 CRC Error: RCU reception support in inconsistent state
Applies to
- Product
- Ventri Ventri Ventri
- Component
- 5265970 (UGP500066) Ventri RCU Assembly Spare Part Replaces UGP500066 5265970 Ventri RCU Assembly Spare Part Replaces UGP500066, see Service Note 43784 *Gantry Gantry Gantry
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- Ventri RCU has an intermittent enable button
- Muga patch FMI 40788 when loaded does not indicate it completed
- How do I capture and print COR and Camera Uniformity (CUA) result screens on Genieacq, then send to a processing station for ACR
- Part number needed for Optima E-Stop
- Collision and E-stop lights are red on the back of the gantry and no movement with Hand Held Controller (HHC) or the genieacq.
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2