Pular para o conteúdo principal
GE Healthcare · MRSigna2236402

CERD DIAGS ASC Static RAM Failure Due to CERD Digital Board.

Symptom

CERD Ram Errors That Do Not Stop Scan or TPS Download Errors do not interrupt scanning. Errors would appear in error log on every tps reset or boot up. Error: 2236402 Host: lx-mr Process: NSP File: TPS:ascd_sram.c Line: 141 CERD DIAGS - ASC Static RAM Failure. ASC Static RAM has failed the data pattern test. Address: 00302670H Data exp: 55555555H Data read: 55555515H

Error message

2236402 CERD DIAGS -ASC Static RAM Failure. ASC Static RAM has failed the data pattern test 2236402 Host: lx-mr Process: NSP File: TPS:ascd_sram.c Line: 141 CERD DIAGS - ASC Static RAM Failure. ASC Static RAM has failed the data pattern test. Address: 00302670H Data exp: 55555555H Data read: 55555515H

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion Universal CERD Digital 2206099 See Long Description This part requires second opinion approval before ordering. See comments in resolution concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues