Millennium VG Error 167 always same position – Melted Plastic Millennium VG Error 167 always same position Whe…

Millennium VG Error 167 always same position – Melted Plastic

Symptom

Millennium VG Error 167 always same position – Melted Plastic Millennium VG Error 167 always same position When rotating millennium VG gantry system always stop with error 167 always on same position. Removing gantry back cover FSE found Melted Plastic on encoder. Contributor 100008553 B.Certo May 7, 2009

Error message

Error #167 (Error 167) MAXIMUM DAC COMMAND. AXIS = 7, ROTATION

Applies to

Models
Millennium
Product
VGH VG and VG Hawkey and Varicam
Component
*Gantry Gantry Gantry
Area
Nuclear Medicine