Millennium VG Error 167 always same position – Melted Plastic Millennium VG Error 167 always same position Whe…
Millennium VG Error 167 always same position – Melted Plastic
Symptom
Millennium VG Error 167 always same position – Melted Plastic Millennium VG Error 167 always same position When rotating millennium VG gantry system always stop with error 167 always on same position. Removing gantry back cover FSE found Melted Plastic on encoder. Contributor 100008553 B.Certo May 7, 2009
Error message
Error #167 (Error 167) MAXIMUM DAC COMMAND. AXIS = 7, ROTATION
Applies to
- Models
- Millennium
- Product
- VGH VG and VG Hawkey and Varicam
- Component
- *Gantry Gantry Gantry
- Area
- Nuclear Medicine