AP Errors Due to PPC Module on AP board.

Symptom

Array processor AP errors during scanning. Scans stop (abort) with an AP error logged. It appears that AP crashes are not related to heavy duty recon jobs or high data acquisition rates. A typical set of stress scans did not reproduce AP crashes. AP error file doesn't reveal any pattern either. Site currently has old style (Reflex-400A) boards. These boards were swapped for the newer Reflex-400B, but there was no change in symptoms. SPR MRIhc06568

Error message

2228010 AP Error

2228016 Internal AP software error See Long Description Please consult file /usr/g/service/log/aperror.log for more information.

2232902 Processor Exception Frame, Bus Error: vector 2, interrupt level 0 Mon Nov 3 13:11:30 1997 Error: 2232902 Host: oc001 Process: NSP File: TPS:excCHECKER Line: 0 Processor Exception Frame, Bus Error: vector 2, interrupt level 0 current pc 0x279bf2, sr 0x3004, task name MC860_CEID_2, fault address 0x400b0008, fault status 0x105 additional information in /usr/g/service/log/vxExLg-4c752740

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion Dual PPC Module 2222532-3 See comments in resolution concept for second opinion requirements before ordering this part See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.