GE Healthcare · CTLightSpeed2601120011233533112260115243
LightSpeed 5.X (60kW)
Symptom
Table would not move in or out. Table would not move in or out
Error message
Error: 260112001 Interference updates are too slow. See Long Description for complete error. Error: 260112001 Interference updates are too slow. Tue Oct 9 06:24:54 2007 Error: 260112001 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr_gt.c Line: 3266 Tue Oct 9 06:24:54 2007 Host : Table Gantry Processor Ermes # : 260112001 Exception Class : Secondary Severity : Pri/Soft File : uif_table_tilt_mgr_gt.c Line# : 3266 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Interference updates are too slow. Table Type: 89 Sides:0x498638b8 states:1233533112 Tilt: 520 mdeg Elevation: 520 um Cradle: 0 um IMS: 0 um Error: 260115243 MTCB cradle encoder data out of range. See long description for details Wed Sep 26 18:21:34 2007 Error: 260115243 MTCB cradle encoder data out of range.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required