No or Little Signal From Coil Elements 5 and 6 of 8CTL Coil due to Wrong Coil Name in Scan Prescription.

Symptom

MRIQ No or Little Signal From Coil Elements 5 and 6 of 8CTL Coil See Long Description and Attachment. This is a 1.5T HDxT system running 15.0 M3 0845.c software. The customer called to complain that the lumbar spine images are missing signal in the coil element 5 and 6 area when using the 8CTL coil. (See the image on the left, below). The customer tried the identical model coil from their other scanner on the same patient and said it worked fine, so the problem must be the coil. See Attachment for the images.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR