Scans Stop Intermittently With Error Code 2248008 Logged Due To Failed Remote RF DIF Board.

Symptom

Scans Stop Intermittently With Error Code 2248008 "RFM EOS semaphore timeout error" See Long Description Scans stop in the middle of the scan with the following error logged. This happens about once per day. (PQR 13096684) 2248008 EXCITE15T NSP APS:DRF 250 RFM EOS semaphore timeout error. IRF did not send enough data to DRF. Please change prescription or perform a TPS Reset. DRF ID: 1 FE ID: 1 segCount: 12 frameCount: 14

Error message

2248008 RFM EOS semaphore timeout error. IRF did not send enough data to DRF.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
5250028 Remote RF Board RF-DIF Remote RF Board 5250028 Remote RF Board 5250028
Area
MR

Full cause & step-by-step resolution