GE Healthcare · MRSigna2248008
Scans Stop Intermittently With Error Code 2248008 Logged Due To Failed Remote RF DIF Board.
Symptom
Scans Stop Intermittently With Error Code 2248008 "RFM EOS semaphore timeout error" See Long Description Scans stop in the middle of the scan with the following error logged. This happens about once per day. (PQR 13096684) 2248008 EXCITE15T NSP APS:DRF 250 RFM EOS semaphore timeout error. IRF did not send enough data to DRF. Please change prescription or perform a TPS Reset. DRF ID: 1 FE ID: 1 segCount: 12 frameCount: 14
Error message
2248008 RFM EOS semaphore timeout error. IRF did not send enough data to DRF.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 5250028 Remote RF Board RF-DIF Remote RF Board 5250028 Remote RF Board 5250028
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.