Symptom

Fastcals are failing with several FRDM errors

Error message

Error: 230007001 Exclusion Test - Row xx:xxx, Channel xxx, Module xx Open Channel* FAILED X-AXIS EXCLUSION CRITERIA Mon Oct 3 11:30:58 2005 Error: 230007001 Host: ctm1 Process: badchanmgr File: BadChannelMap.cxx Line: 544 Exclusion Test - Row xx:xxx, Channel xxx, Module xx Open Channel* FAILED X-AXIS EXCLUSION CRITERIA *.

Error: 230007002 Exclusion Test - Row xx:xxx, Channel xxx, Module xx Open Channel* FAILED PR OXIMITY CRITERIA * See LD for a sample of entire error message. Fri Oct 28 09:32:03 2005 Error: 230007002 Host: presbyplano Process: badchanmgr File: BadChannelMap.cxx Line: 638 Exclusion Test - Row 64:32B, Channel 695, Module 44 Open Channel* FAILED PR OXIMITY CRITERIA *.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5197802 Cherokee 5kW 48vdc power supply with fuse board. USED ON VCT gantry rotating frame. Supercedes previous Pioneer part numbers 5197802, 5183361, 5154556, and 5126259 Part #: 5197802 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCTHD 48V PS with FRU packaging FRU: Yes Status: Orderable Comments: **FE ASSIST** 5197802-2 is used on a HD-CT750 system, Do not use on a VCT system.(212001261 9/30/2010) .Per HDVCT RSPL, 212008898, 11/6/08 Part #: 5197802 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT Pioneer 5kW Power Supply Assembly Collector, with FRU packaging material FRU: Yes Status: Orderable Comments: Per ECR 2037621, 212008898, 1/5/07. If no stock, please order part # 5183361 --Description: VCT 5kW Power Supply Assembly, with ferrite and external shield. Part number change up history information: P/N 5197802 Cherokee " supply has two main differences: 1) The newer Cherokee supply is approximately 1.2 kg (2.6 lbs) Heavier then the older Pioneer Supply - this may cause a Gantry rebalance. 2) The large Ferrite bead is not visible on the Cherokee Supply because the ferrite is internal on the Cherokee Supply. P/N 5183361 was created to add an external shield that diverts the magnetic field energy away from the low channel DAS/Detector channels. The low channels on the VCT DAS/Detector are sensitive to magnetic fields that are produced from the power supply. This p/n replaced 5154556. P/N 5154556 was created to add a ferrite core to the 48 VDC output cable of the supply. This change was made to address streaking imaging artifacts that were sometimes seen with low mA scans by decreasing noise induced into the DAS/Detector. This p/n replaced 5126259. P/N 5126259 was the original VCT gantry rotating frame 48 vdc power supply. 5183361 48 vdc power supply with ferrite core and external shield 5154556 48 vdc power supply with ferrite core 5126259 48 vdc power supply without ferrite core or magnetic shield
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.