The Positioner vertical lift intermittently does not hold its position.
Symptom
The Positioner vertical lift intermittently does not hold its position. See Long Description The D5K positioner vertical lift does not hold position from lift motor drive stop to brake engagement. It will intermittently/typically drop about an centimeter, post vertical travel request (at the travel motion stopping point) both in up and down motion. The vertical brake does engage properly, but does have a slight delay (fraction of a second) in actuation, which is enough time to allow noticeable drift.