Ventri System does not pull the table out when UNLOAD key is pushed.

Symptom

Ventri System does not pull the table out when UNLOAD key is pushed. The table will move down to 76.0 cm but does not proceed to move out of the ring System does not pull the table out when UNLOAD key is pushed. The table will move down to 76.0 cm but does not proceed to move out of the ring. Can hear the linear brake release but linear axis never moves. Brake will engage again. Briefly displays UNLOAD CANCELLED message on the monitor. At this time the table can not be moved out with the RCU either - brake releases but no motion. IF the table is raised a small amount with the RCU to f.ex. 76.1 or 76.3 cm then linear moves with RCU IN/OUT (UNLOAD still doesn't work as it takes vertical back to 76.0 cm). The same issue intermittently happens when the table is positioned using automatic load (SET). It started as intermittent issue but now it happens more like on 60-70 % of all LOADS / UNLOADS CSO #1-1797576527

Error message

ERROR_GCU_FAILED_RPDO_EMCY (Error #1369)

ERROR_GCU_AXIS_TRACKING_EMERGENCY (Error 1345)

Applies to

Product
Ventri Ventri Ventri
Component
CBL500067 Ventri Table longitude motion power cable CBL500067 Ventri Table longitude motion power cable *Patient Table *Patient Table *Patient Table
Area
Nuclear Medicine

Full cause & step-by-step resolution