GE Healthcare · CTLightSpeedVCT184516
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2
Symptom
Fastcals failed to complete. Fastcals failed to complete
Error message
Error: 184516 The CPU was reset by the CPU watchdog. See long description for full error Mon Apr 13 07:23:51 2009 Error: 184516 Host: STC Process: Control File: AXitest.c 1.14 Line: 116 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:23:45:285 Log Series:2 The CPU was reset by the CPU watchdog. INFO: Watchdog timeout may have been caused by either a hardware or software error. No further analysis is possible.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT