GE Healthcare · MRSigna2241545
Will Not Scan with Gradient Errors.
Defective GP Module.
Symptom
Will Not Scan with Gradient Errors.
Error message
2241545 The GIP reported a second Long Overload on the X Axis of the GPM. Tue Apr 20 20:24:34 2004 Error: 2241545 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 653 The GIP reported a second Long Overload on the X Axis of the GPM. Power cycle the main breaker on GPM Power Supply to recover from this condition. If this fault occurs again, swap the GPM axis. Run diagnostics.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR