Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2300160731941459962523000650611433230016059

LightSpeed 4.X - Octane

Symptom

Image reconstruction failures. Image reconstruction failures

Error message

Error: 230016073 rc:-5 Post main failed See long description Tue Mar 31 12:27:33 2009 Error: 230016073 Host: ice-post Process: tsdc1 File: sdc post main.c Line: 982 Tower 1, Dsp id 1 Function Image Generation : Preprocessing Current State :0xffffffff Time Stamp : 19414 sec : 599625 nsec PATH : sdc post main.c:982/sdc post apps.c:950/vw view loop.c:292 /sdc post main .c:1009 INFO : vw loop nslice overlay rc:-5 Post main failed

Error: 230006506 Reconmgr received job_status error RECON_JOB_FAILED_RETRY from SDC Control See LD for complete error Sat Jul 1 19:39:27 2006 Error: 230006506 Host: ice Process: reconmgr File: rm_process_job.c Line: 491 Function: Image Generation : SBC Processing Job Type: Prospective Scout (s Scan: 11433/1/2 Image: 11433/1/2 Recon Seq Id: 2824 Reconmgr received job_status error RECON_JOB_FAILED_RETRY from SDC Control in the sdc done event.

Error: 230016059 dma status wait failed rc:104 Start of input dma failed See long description Tue Mar 31 12:27:50 2009 Error: 230016059 Host: ice-post Process: tsdc1 File: sdc post apps.c Line: 950 Tower 1, Dsp id 1 Function Image Generation : Preprocessing Current State :0xffffffff Time Stamp : 19430 sec : 850634112 nsec

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2300482 Pegasus / Nancy IG Board for ICE Box in LS 4x only See long description 2300482 Pegasus / Nancy IG Board for ICE Box in LS 4x only Part #: 2300482 Part Type: GP REPAIRABLES Part Weight: 0.58 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: IG Bd. PEGASUS / NANCY ASSEMBLY FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues