Unable to Perform Grafidy3 with New Grafidy3 6-Coil Kits Received March and April 2009 due to 1.5T Grafidy3 6-Coil Kit 5…

Unable to Perform Grafidy3 with New Grafidy3 6-Coil Kits Received March and April 2009 due to 1.5T Grafidy3 6-Coil Kit 5135527 has Incorrectly Programmed Coilid Code.

Symptom

Unable to Perform Grafidy3 with New Grafidy3 6-Coil Kits Received March and April 2009. See Long Description Unable to perform Grafidy3 with new Grafidy3 6-Coil kits (officially known as "6-Coil ECMT Kit, 1.5T") received March and April 2009. When attaching the Grafidy3 6-Coil kit, it results in a pop-up asking for installation of coil configuration CD. No error is logged.

Error message

Popup Message "Install Coil Configuration CD" See Long Description A pop-up message appears to the operator asking for installation of coil configuration CD. No error is logged.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
5135527 1.5T Grafidy3 6-Coil Kit (6-Coil ECMT Kit, 1.5T) Officially known as "6-Coil ECMT Kit, 1.5T"
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.