Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT182100381431821251598546159835012320923

LightSpeed 4.X - GRE

Symptom

Scanner shuts down on its own during scanning. Scanner shuts down on its own during scanning

Error message

Error: 182100 HW/SW Error: Did not receive message from local exception handler. Will need a Hardware Reset. See long description for full error Fri May 30 08:34:13 2008 Error: 182100 Host: STC Process: Master Exc Handler File: exm_verify_wdog.c Line: 202 Function: No System Function Reported Scan: 38143/201/0 Type: None/Unknown Exception Level: Primary Time: 08:34:13:185 Log Series: 486 Notify Verification TIMEOUT from uuid 292 (292=STC; 92=ETC). HW/SW Error: Did not receive message from local exception handler. Will need a Hardware Reset.

Error: 182125 LAN Watchdog TIMEOUT with OBC or ETC (uuid 93, 193) (SW opening abort relay). See LD for full errors Mon Dec 6 22:19:33 2004 Error: 182125 Suite: CT01 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Method: No Method Line: 773 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 22:19:32:730 Log Series: 27 LAN Watchdog TIMEOUT with OBC (uuid 193) (SW opening abort relay). HW/SW Error: Did not receive LAN Watchdog message from OBC. (Last Wdog Probe Sent:1598546 Last Response Rcvd:1598350 (in Ticks)). Error could be for the ETC also Reference Thu Jan 12 09:54:50 2006 

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required