GE Healthcare · MRSigna
CWA ID132592 Banding and Signal Drop in CUBE is due to Phase Correction Choosing Echoes of Lower SNR in Cube.
Symptom
MRIQ Image Banding Artifacts and Signal Drop on CUBE Sequences See Long Description and Attachment Banding artifacts and signal drop on CUBE sequences, no errors reported. See attachment for example images of artifacts.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required