Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260112014833869

LightSpeed 5.X (80/100kW)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260112014 Touch sensor ? Touch Sensor was PRESSED See Long Description for full error Error: 260112014 Touch sensor ? Touch Sensor was Pressed. Sun Mar 12 15:14:25 2006 Error: 260112014 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr_gt.c Line: 9013 Sun Mar 12 15:14:24 2006 Host : Table Gantry Processor Ermes # : 260112014 Exception Class : Secondary Severity : Sec/Soft File : uif_table_tilt_mgr_gt.c Line# : 9013 Function : PATIENT ANATOMY POSITIONING : Tilt Drive Scan Type : None/Unknown Scan: 0/0/0 Touch sensor BACK Touch Sensor was PRESSED. Elevation position where touch Sensor was Pressed:0 um. Tilt position where touch sensor was pressed:0 mdeg. Remote Tilt button is DISABLED. Use local gantry controls to tilt the gantry. Or Mon Apr 9 09:29:44 2007 Error: 260112014 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr.c Line: 9990 Mon Apr 9 09:29:44 2007 Host : Table Gantry Processor Ermes # : 260112014 Exception Class : Secondary Severity : Pri/Soft File : uif_table_tilt_mgr.c Line# : 9990 Function : PATIENT ANATOMY POSITIONING : Elevation Drive Scan Type : None/Unknown Scan: 317/6/0 Touch sensor FORWARD Touch Sensor was PRESSED. Elevation position where touch 

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2277405 Front / Small Gantry Tilt Forward Sensor / Interference Switch used on H2 gantry See Long Description for Details 2277405 Front / Small Gantry Tilt FORWARD Sensor / Interference Switch used on H2 Part #: 2277405 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PATIENT SENSOR, SMALL FRU: Yes Status: Orderable Comments : FLS - SAFETY TILT INTERFERENCE SWITCH FRONT USED ON H2 COMMON TO MFG. MS26561 DSM 1/7/01 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues