LightSpeed 5.X (80/100kW)
Symptom
Scan hardware stopped scan with DAS Board errors. Scan hardware stopped scan DAS Board errors Scanner Hardware Stopped Scan message appears in the operator console area during Axial Scan data acquisition with Error: 260001023 The DCB has notified the system of an error condition which prevents scanning and error: 260007402-Converter Board(s) Asserted a Board FAULT cards 76,78,80,82,84,86,88,90,92,94,96 and Error #260007774 list of converter cards from previous error message Problem was intermittent Then after time the errors would occur more often and eventually the problem was consistent [insite@erct ~]$ /usr/g/insite/bin/viewlog Sun Aug 29 Error: 260001023 Host: OBC Process: HOP File: hop erlog.c Line: 367 Function: No System Function Reported Scan: 50000/1/2 Type: Axial Exception Level: Pri/Most Severe Time: 08:37:10:145 Log Series: 120 The DCB has notified the system of an error condition which prevents scanning. See log for further information:^C Error message code: 260002358 Notification level: 5 Sequence number: 11
Error message
Error: 260101028 The DCB has shutdown due to an error. A controller reset is required to continue. 1095852103 0 1 Wed Sep 22 11:21:43 2004 260101028 4 orp orp_subsys rcib_erlog.c 217 Wed Sep 22 11:21:41 2004 Host : On-Board Rotating Processor Ermes # : 260101028 Exception Class : Soft Severity : Pri/Soft File : rcib_erlog.c Line# : 217 Function : No System Function Reported Scan Type : None/Unknown Scan: 26/2/0 The DCB has shutdown due to an error. A controller reset is required to continue. 1095852103 0 1 Wed Sep 22 11:21:43 2004 260101028 4 orp orp_subsys rcib_erlog.c 217 Wed Sep 22 11:21:41 2004 Host : On-Board Rotating Processor Ermes # : 260101028 Exception Class : Soft Severity : Pri/Soft File : rcib_erlog.c Line# : 217 Function : No System Function Reported Scan Type : None/Unknown Scan: 26/2/0 The DCB has shutdown due to an error. A controller reset is required to continue. Error: 260007483 DCB detected Coverter Board Fault Line Asserted. The Converter Boards Are Not In A Valid Configuration. Converter Board Fault Line Will Be Disabled Until Next Das Reset. See LD for full error Sun Jul 31 12:39:26 2005 Error: 260007483 Host: orp Process: orp_subsys File: brdflt.c Line: 19
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2287134 Mdas or 2287134-4 GDAS Jumper Cable Right or Left, connects to Center Chassis See Long Description for IMPORTANT differences between these two part numbers ! ! 2287134 MDAS or 2287134-4 GDAS Jumper Cable Right or Left side of GDAS that interconnects with the Center chassis NOTE: Check your DAS part number before ordering! 2287134 THIS PART IS FOR MDAS 2358139 ONLY 2287134-4 THIS PART IS FOR GDAS 2358139-2 ONLY Part #: 2287134-4 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Harness, GDAS Jumper, ED.2 EMC FRU: Yes Status: Orderable Comments: PART ADDED PER VCT RSPL AND IS EMC2 COMPLETE 12/01/2004 Part #: 2287134 Part Type: GP CONSUMABLES Part Weight: 0.66 LBR Repairable: No Returnable: No Hazardous Material: N Description: MDAS JUMPER HARNESS FRU: Yes Status: Orderable Comments: FLS - MDAS JUMPER HARNESS - RT/LT ADDED VIA PCN 220296. VENDOR CHANGED FROM NORTECH TO SANMINA SCI IN MEXICO PER E-NPI TRANSFER "SANMINA CABLE TRANSFER". MS20544 ON 6/7/02.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required