Pular para o conteúdo principal
GE Healthcare · CTLightSpeed

LightSpeed 5.X (60kW)

Symptom

Image recon is failing intermittently - document in RFS - critical issue See long description for details. Image recon is failing intermittently.

Error message

ig1 kernel: rac: PCI ERROR: Detected Parity Error See long description for details ig1 kernel: rac: PCI ERROR: Detected Parity Error Use commands below to search for Errors in the IG messages log: rsh darc rsh ig1 (this could be ig1, ig2, or ig3) [root@ig1 ~]# cd /var/log [root@ig1 log]# cat messages | grep Parity Feb 18 14:41:37 ig1 kernel: rac: PCI ERROR: Detected Parity Error Feb 18 14:41:40 ig1 kernel: rac: PCI ERROR: Detected Parity Error Mar 1 09:44:31 ig1 kernel: rac: PCI ERROR: Master Data Parity Error Mar 1 09:44:31 ig1 kernel: rac: PCI ERROR: Detected Parity Error [root@ig1 ~]# cd /var/log [root@ig1 log]# cat messages | grep ERROR Mar 1 09:44:31 ig1 kernel: rac: PCI ERROR: Master Data Parity Error Mar 2 16:05:07 ig1 kernel: rac: PCI ERROR: Signaled Target Abort Mar 2 16:05:52 ig1 kernel: rac: PCI ERROR: Signaled Target Abort

ig1 kernel: rac: PCI ERROR: Master Data Parity Error See LD for details ig1 kernel: rac: PCI ERROR: Master Data Parity Error rsh darc rsh ig1 ( ig1, ig2, or ig3 ) cd /var/log tail -1000 messages | more Look for any of the IG errors listed below: Aug 29 15:20:24 ig1 kernel: rac: PCI ERROR: Master Data Parity Error Aug 29 15:20:24 ig1 kernel: rac: PCI 

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2362872 Westville IG Module used on GOC3 and GOC4 consoles. See Long Description for details. 2362872 Westville IG Module used on GOC3 and GOC4 consoles. Part #: 2362872 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 19.84 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: IMAGE GENERATION COMPUTER - GRE FRU: Yes Status: Orderable Comments: **FE Assist** applies to CT Not Pet systems. If no stock is available order a Jarrell IG 5147443 to replace a Westville IG (2362872), you MUST order p/n 5193989 software release 06MW29.7, and perform a LFC. See Service note SN06101010A. PET must receive the Westville IG.(12/12/06 212001261) *** 2ND OPINION NOTES *** See the following URL for additional command line diagnostic troubleshooting: http://gemsvst4p.med.ge.com:7005/facets/CT/GRE%20command%20line%20diag%20test.pdf?
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.