Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2600010075004226010800050052260004000

LightSpeed 5.X (80/100kW)

Symptom

Scanning aborted midway through exam with scan communication errors.

Error message

Error: 260001007 CAN serial link error. A CC error was detected when sending a message to the DCB See LD for more details Wed Nov 22 05:44:35 2006 Error: 260001007 Host: OBC Process: HOP File: hop_cc_send.c Line: 196 Function: No System Function Reported Scan: 50042/6/0 Type: None/Unknown Exception Level: Pri/Soft Time: 05:44:35:070 Log Series: 80 CAN serial link error. An CC error was detected when sending a message to the DCB. Message code: 1004 Node ID: 51 Object ID: 0 Send status: 5

Error: 260108000 Collimator firmware object: Lost communications with controller GCAN error sending message = 5. See Long Description. Thu Dec 22 16:15:57 2005 Error: 260108000 Host: orp Process: orp_subsys File: ss_comm.c Line: 497 Thu Dec 22 16:15:55 2005 Host : On-Board Rotating Processor Ermes # : 260108000 Exception Class : Abort Severity : Primary File : ss_comm.c Line# : 497 Function : SYSTEM COMMUNICATIONS : Rotating Controller I/F Bus Scan Type : None/Unknown Scan: 50052/1/0 Collimator firmware object: Lost communications with controller GCAN error sending message = 5.

Error: 260004000 DAS firmware object: Lost communications with controller See Long Description for more details Wed Nov 2

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2361853-2 GDAS DCB 16-SLICE replaces 2361853 See LD for more Info 2361853-2 GDAS DCB 16-SLICE replaces 2361853 Part #: 2361853-2 Part Type: GP CONSUMABLES Part Weight: -59999.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: GDAS DCB FRU: Yes Status: Orderable FLS - GDAS DCB (16-SLICE) USED ON: LIGHTSPEED PRO 16 GDAS ADDED PER GDAS RSPL MS13452 This board has been replaced by 2361853-2, FE NOTE: Site MUST have M4FMI completed before this board will function properly, ms13452
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.