What is the Meaning of Error Code 2236727 Analog Receiver Anti Alias Filter Test RESULTS.
Symptom
What is the Meaning of Error Code 2236727 Analog Receiver Anti Alias Filter Test RESULTS See Long Description The FE ran CERD Receiver diagnostics and received the errors shown below. He is unsure about the meaning of error code 2236727 "CERD Diags : Analog Receiver Anti Alias Filter Test RESULTS". Is this a real problem? What is the meaning of this error and what should the FE do? Wed Mar 25 13:46:24 2009 Error: 2238002 Host: LXMR Process: diags File: diagsMain.cc Line: 1387 Starting CERD Receiver diags. (command word : 00000100 loop count : 1) Wed Mar 25 13:46:25 2009 Error: 2236727 Host: LXMR Process: diags File: diagsMain.cc Line: 2332 CERD Diags : Analog Receiver Anti Alias Filter Test RESULTSReceiver channel 2 has bandwidth of 100 kHz. This is a configuration message, NOT a test error. Wed Mar 25 13:46:25 2009 Error: 2236727 Host: LXMR Process: diags File: diagsMain.cc Line: 2332 CERD Diags : Analog Receiver Anti Alias Filter Test RESULTS Receiver channel 3 has bandwidth of 100 kHz. This is a configuration message, NOT a test error. Wed Mar 25 13:46:25 2009 Error: 2236727 Host: LXMR Process: diags File: diagsMain.cc Line: 2332 CERD Diags : Analog Receiver Anti Alias Filter Test RESULTS Receiver channel 4 has bandwidth of 100 kHz. This is a configuration message, NOT a test error. Wed Mar 25 13:46:25 2009 Error: 2238004 Host: LXMR Process: diags File: diagsMain.cc Line: 2191 Diags completed.
Error message
2236727 CERD Diags Analog Receiver Anti Alias Filter Test RESULTS Receiver channel X has bandwidth of 100 kHz See Long Description Error: 2236727 Host: LXMR Process: diags File: diagsMain.cc Line: 2332 CERD Diags : Analog Receiver Anti Alias Filter Test RESULTS Receiver channel X has bandwidth of 100 kHz. This is a configuration message, NOT a test error. Note: the variable X shown here will actually be the Receiver channel 1, 2, 3, or 4.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.