GE Healthcare · CTLightSpeedVCT18200450238431109430857182026
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
Customer gets Error message 'DAS gain calibration has not been run since converter boards were changed' while running fast cals. Customer gets Error message 'DAS gain calibration has not been run since convertor boards were changed' while running fast cals.
Error message
Error: 182004 LAN Watchdog TIMEOUT See LD for the full error code Mon Feb 14 10:28:30 2005 Error: 182004 Suite: SJ01 Host: ETC Process: Local Exc Handler File: exl_lan_in_mail.c Method: No Method Line: 175 Function: No System Function Reported Scan: 50238/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 10:28:30:550 Log Series: 31 LAN Watchdog TIMEOUT. HW/SW Error: Did not receive LAN Watchdog message from STC. (Current Time:431109 Last Wdog LAN Probe Time:430857 (in Ticks)) Error: 182026 Did not receive LAN Watchdog message from STC See long description for full error Mon Mar 23 08:32:18 2009 Error: 182026 Host: ETC Process: Local Exc Handler File: exl_lan_in_mail.c Line: 210 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 3447 Log Series: 3 LAN Watchdog TIMEOUT. HW/SW Error: Did not receive LAN Watchdog message from STC. Putting Hardware Into Safe State. (Current State:3447)
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution