MRIQ Ghosting and MRSPT FSE Phase Offset Off Isocenter Due to HOS not terminated.

Symptom

MRIQ Ghosting and MRSPT FSE Phase Offset Off Isocenter Customer complaint of little ghosting on some images, SPT failed on FSE (severe on Zoom Head, phase offset noisy out of isocenter on the 3 plans), due to HOS termination cable missing

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR