Pular para o conteúdo principal
GE Healthcare · MRSigna10033

Cannot Archive Due to MOD drive.

Symptom

MOD Drive will not attach for archive. The system is a 5X Horizon system that was upgraded to use the Maxoptix MOD drive and media. An FE tried to order the Maxoptix T5-2600 drive thru Cares, but due to end of life issues, the drive changed up to a Sony MOD drive. After installing the new Sony drive, the system would recognize the drive as a Sony but could not attach or archive using the drive.

Error message

10033 There is no media in the drive Mon Apr 26 13:46:16 2004 Error: 10033 Host: MR-MN Process: arsjob File: ArchiveDevice.c Line: 4780 There is no media in the drive.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues