GE Healthcare · MRSigna

For HDxt (15x), Fluoro Trigger and PURE option are not compatible.

Symptom

Fluoro trigger scan is started, the system scans, but images are never displayed. The GUI buttons are grayed out and an hourglass is displayed for a long time which prevents the operator from having any control over the sequence. Eventually, the hourglass will go away and the patient/protocol info is displayed in the image display window but never an image displayed.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR