GE Healthcare · MRSigna22488212248850
RF Amplifier Fault 44 Due to SRFD2.
Symptom
Auto Prescan failed, RF Amplifier is not ready Getting another message that RF Amplifier is powering but message never goes away.
Error message
2248821 Analogic RF Amplifier 1 Device Fault 44. RF FET device is shorted. Replace amplifier Mon Nov 27 10:03:08 2006 Error: 2248821 Host: VDMRI Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2218 Analogic RF Amplifier 1 Device Fault 44. RF FET device is shorted. Replace amplifier.-> 2248850 RF Amplifier fault 71 Mon Nov 27 10:10:04 2006 Error: 2248850 Host: VDMRI Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2218 RF Amplifier fault 71.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.