GE Healthcare · CTLightSpeedBrightSpeedHiSpeed26011827350357
LightSpeed 5.X (80/100kW)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118273 (110-0805) Inverter HW Temperature too high. See LD for details Mon Jan 8 06:44:06 2007 Error: 260118273 Host: orp Process: orp_subsys File: gen_exec.c Line: 2376 Mon Jan 8 06:44:05 2007 Host : Jedi Ermes # : 260118273 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : None/Unknown Scan: 50357/1/0 (110-0805)Inverter HW Temperature too high. Inverter heatsink surface has reached 67 degrees The above error seem to be caused by repeated tube spitting and arcing.
Applies to
- Models
- LightSpeed, BrightSpeed, HiSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2281950-3 or 2281950-4 : JH4 INVERTER WITH SHIELDS used on LS 5.X (80/100), LS 7.X See Long Description for Details 2281950-3 Can also use 2281950-4 JH4 INVERTER WITH SHIELDS used on LS 5.X (80/100), LS 7.X Part #: 2281950- 3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: JH4 INVERTER WITH SHIELDS FRU: Yes Status: Orderable Comments: added per ecr 2005924, as replacement for 2281950 and 2281950-2. ms13452 040923 Set as Critical Part for Product CT_VCT, ms04278, 05Oct14 - If you receive an inverter as FRU (P/n 2281950, 2281950-2, 2281950-3) that must be installed to replace a defective inverter on site, check first Part number of PPC kV control board in this FRU: - if it is labelled 2351200-11 rev 001 this board needs to be replaced with the kV control board that is installed in the defective inverter: PPC board replacement, please refer to PPC FRU replacement procedure in service manual. - if it is not labelled 2351200-11 rev 001, the board installed on FRU is good: no modification is needed. Shawn Fox 3-24-05 Part #: 2281950 -4 Part Type: GP CONSUMABLES Part Weight: 0.00 LBRPart #: 2281950-4 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: JH4 INVERTER WITH SHIELDS FRU: Yes Status: Orderable Comments: Added per ECR 2018129, THIS PART IS BACKWARD COMPATIBLE REPLACEMENT FOR 2281950-3, -2, & NO DASH. MS13452 050919 Repairable: No Returnable: No Hazardous Material: N Description: JH4 INVERTER WITH SHIELDS FRU: Yes Status: Orderable Comments: Added per ECR 2018129, THIS PART IS BACKWARD COMPATIBLE REPLACEMENT FOR 2281950-3, -2, & NO DASH. MS13452 050919 2266521-5 HV Tank (replaces 2266521-4 and 2266521-3 - backward compatable) used on LS 7.X, LS 5.X and LightSpeed RT. See Long Description for details 2266521-5 HV Tank (replaces 2266521-4 and 2266521-3 - backward compatable) used on LS 7.X, LS 5.X and LightSpeed RT. Dash Tanks includes the EMI Shielding needed for G-Das Part #: 2266521-5 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Part #: 2266521-4 Part Type: GP CONSUMABLES Part Weight: -5499945.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Part #: 2266521-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 200.84 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Comments: added per ecr 2005924, as replacement for 2266521 and 266521-2. ms13452 040923 **If there is no Local Stock for the -3 part, you may order the -4 version which is backward compatible** ms0262 060815 *** 2ND OPINION NOTES *** Run the No-load KV test to verify HV tank is the error.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.