Pular para o conteúdo principal
GE Healthcare · CTLightSpeed182407185122

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Table will not move. Table will not move

Error message

Error: 182407 Velocity error detected when trying to find the Index Pulse Sun Aug 29 13:25:59 2004 Error: 182407 Host: ETC Process: Elevation File: CmdMotion.c 1.35 12/23/93 Line: 174 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 13:25:59:690 Log Series: 34 HW/SW error: Velocity error detected when trying to find the Index Pulse. Hardware Inhibit Switches: 0xc0d8 Elev Direction: 0 (0 = Up 1 = Down). INFO: Probable Problem could be 180V Power is LOW.

Error: 185122 24V Power Supply Out Of Tolerance Sun Sep 5 16:04:34 2004 Error: 185122 Suite: RP21 Host: ETC Process: Control File: etcPowSupTest.c 1.4 2/16/99 Method: No Method Line:117 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 3588 Log Series: 2 HW error: 24V Power Supply Out Of Tolerance: A/D reading = 0x4 Calc Voltage = 0.03 (Volts) Tolerance = 3 (Percentage) INFO: If the supply is low the problem may be that the Xray Drive Power is OFF

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.