Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1111710008

LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry

Symptom

Technologist CANNOT remove Exams from the system browser. Unable to remove images until Reboot is completed

Error message

Error: 11117 Remote does not support push for SOP_Class_UID See Long Description Two version of this error Exist. “DICOM Enhanced SR SOP” class. UID 1.2.840.10008.5.1.4.1.1 .88.22 “DICOM X-ray Radiation Dose SR SOP” class UID 1.2.840.10008.5.1.4.1.1 . 88.67 Fri Feb 13 15:22:35 2009 Error: 11117 Host: msec8x1 Process: dcserver File: dcspush.c Line: 2456 Remote does not support push for SOP_Class_UID 1.2.840.10008.5.1.4.1.1.88.67 , so all images with this SOP_class_UID have been rejected Fri Feb 13 15:22:35 2009 Error: 11117 Host: msec8x1 Process: dcserver File: dcspush.c Line: 2456 Remote does not support push for SOP_Class_UID 1.2.840.10008.5.1.4.1.1.88.22 , so all images with this SOP_class_UID have been rejected “DICOM Enhanced SR SOP” class. UID 1.2.840.10008.5.1.4.1.1 .88.22 “DICOM X-ray Radiation Dose SR SOP” class UID 1.2.840.10008.5.1.4.1.1 . 88.67 * SR stand for "Structured Report"

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
08BW44.1 Software for all BrightSpeed Systems with AIO Console and all RT16 Systems with GOC4 Console See Long Description 08BW44.1 Software is used on all BrightSpeed Systems with AIO Consoles and all RT16 Systems with GOC4 Consoles. Software collector called “08BW44.X” p/n 5328867 contains the following items: 5323837 - LightSpeed Xtream Console Operating System Version - GEMS Linux 6.3.8 5323838 - LightSpeed Applications Software DVD Version 08BW44.X 5328828 - Xtream Serial-Over-LAN Service Software Version - 2.0 Supports: Westville or Jarrell DARC/VDARC 5323839 - Class C 5323840 - Class M NOTE: Class C/M install CD are not shipped with system. NOTE: The Xtream Serial-Over-Lan Service Software is not used during the LFC unless instructed via a pop-up message box "SOL Service - ...run the Serial Over LAN Service CD...". This CD is loaded directly into the DARC Node. Then press the reset (or cycle power) at the front panel of the DARC Node as required. This CD will auto-eject when the SOL load process is complete. NOTE: The Xtream Serial-Over-Lan CD is not used for All-In-One Console.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.