Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2601151902601151917156429999

LightSpeed 5.X (60kW)

Symptom

Cradle will not move into the gantry; motor is heard running and only occurs when cradle is home and latched. Cradle will not move into the gantry; motor is heard running and only occurs when cradle is home and latched with Error: 260112011 Received Table Rx but the Table was NOT READY for Scanning and Error: 260115191 Cradle position is out of tolerance

Error message

Error: 260115190 Cradle encoder fault See the long description for the full error Wed May 25 17:35:00 2005 Error: 260115190 Host: tgp Process: tgp_subsys File: mtcb_cmc_isrfunc.c Line: 163 Wed May 25 17:35:00 2005 Host : Table Sub-System Ermes # : 260115190 Exception Class : Soft Severity : Pri/Soft File : mtcb_cmc_isrfunc.c Line# : 163 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle encoder fault .

Error: 260115191 Cradle position is out of tolerance See long description for full error Thu Sep 1 09:04:12 2005 Error: 260115191 Host: tgp Process: tgp_subsys File: mtcb_cradle_monitor.c Line: 414 Thu Sep 1 09:04:12 2005 Host : Table Sub-System Ermes # : 260115191 Exception Class : Soft Severity : Pri/Soft File : mtcb_cradle_monitor.c Line# : 414 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle position is out of tolerance. cradle_position(um) : 1666 target_position(um) : 71564 velocity (um/s) : 29999 fpga status : 0x8140

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues