Getting GPM errors on all 3-axis due to X-Axis MiniGRAM Failure.

Symptom

Getting GPM errors on all 3-axis. Customer cannot scan and getting GPM errors on all 3-axis indicating power was off.

Error message

2241508 The GIP is reporting ? Axis GPM Power Off Wed Jun 30 07:42:27 2004 Error: 2241508 Host: SJHMR Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 303 The GIP is reporting X Axis GPM Power Off. If all 3 axes are reporting this fault, check the GPM Power Supply circuit breakers and cables. Check GPM fuses and the cable between the GIP and the GPM.

2241510 The GIP is reporting GPM Power Supply Power Off. Mon Jun 25 23:08:14 2001 Error: 2241510 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP Line: 1882 The GIP is reporting GPM Power Supply Power Off. Check the GPM Power Supply circuit breakers and incoming power. Check the cable between the GIP and the GPM Power Supply.

2241549 The GIP reported several miniGRAM Power Supply Faults. After the miniGRAM Power Supply fault LED turns off, power cycle main breaker on Power Supply to recover. If this fault occurs again, check GRAM Tuning, the miniGRAM, miniGRAM Power Supply. 2241549 Host: lx-mr Process: NSP File: HPC PORT ID:GAP/GIP/GP Line: 2394 The GIP reported several miniGRAM Power Supply Faults. After the miniGRAM Power Supply fault LED turns off, power cycle main breaker on Power Supply to recover. If this fault occurs again, c

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2169727-4 or T3503TT Mini GRAM Assembly
Area
MR