Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT200350195178628178517

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 200350195 Unexpected message descriptor received when opening a UUID XXX See the LD for the full error codes Error: 200350195 Unexpected message descriptor received when opening a UUID XXX Mon May 10 16:44:42 2004 Error: 200350195 Host: ct01 Process: coordinator File: new_coord_main.c Line: 3744 Running Version Management Tool Function: System Communications : Control Network Test: Version Management Tool Test status: starting Unexpected message descriptor received when opening a UUID mail box. UUID: 160 (OBC HOP) Message descriptor: 2122 Sun Oct 29 15:46:48 2006 Error: 200350195 Host: ct02 Process: coordinator File: new_coord_main.c Line: 3749 Running Version Management Tool Function: System Communications : Control Network Test: Version Management Tool Test status: starting Unexpected message descriptor received when opening a UUID mailbox. UUID: 220 (STC Axial) Message descriptor: 2122

Error: 178628 The Request To Query the Flash Version Failed. Controller: (varies) See Long Description for complete error and examples. Error: 178628 The Request To Query the Flash Version Failed. Controller: (varies) Mon May 10 16:44:42 2004 Error: 178628 Host: ct01 Process: vertoolTA Fil

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2231813 OBC rear panel 2231813 OBC rear panel 2231813 OBC rear panel
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.