Pular para o conteúdo principal
GE Healthcare · MRSigna221853722333962234205

Frame Data Buffer and CERD Errors Due to IPG.

Symptom

Unable to Scan with CERD, RAP Time Out and Frame Data Buffer Errors See Long Description 2234205 The RAP timed out during acquisition. Internal software or hardware error, please run CERD diagnostics! 2233396 The Frame Data Buffer Queue has overflowed. May need a longer TR time, or more time between echoes. Please restart scan. 2218537 Scan stopped, by Cerd Error Server request.

Error message

2218537 Scan Stopped, by Cerd Error Server Request Fri Jan 18 16:01:10 2002 Error: 2218537 Host: lx-mr Process: scn File: MsaScanMgr.m Line: 2487 Scan stopped, by Cerd Error Server request.

2233396 The Frame Data Buffer Queue has overflowed. May need a longer TR time, or more time between echoes. Please restart scan. Thu Feb 22 17:59:07 2007 Error: 2233396 Host: MENEMR Process: NSP File: TPS:cap: c_int45 Line: 4 The Frame Data Buffer Queue has overflowed. May need a longer TR time, or more time between echoes. Please restart scan.

2234205 The RAP timed out during acquisition. Internal software or hardware error, please run CERD diagnostics! Thu Feb 22 17:59:07 2007 Error: 2234205 Host: MENEMR Process: NSP File: TPS:asc_ba_isr Line: 1 The RAP timed out during acquisition. Internal software or hardware error, please run CERD diagnostics!

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2160818-7 IPGII Board SUN STYLE FRONT PANEL See Long Description IPG2 Board 2160818-7 FEMC Part Description: IPGII SUN STYLE FRONT PANEL
Area
MR

Full cause & step-by-step resolution