Lightspeed 1.X - Octane
Symptom
Root Partition is almost full ( 98%) - customer is unaware of this condition and store log does not help. See Long Description for more info VCT Root Partition is almost full ( 98%) -Customer is unaware of this condition and store log does not help. Once the Root partion is full the System will reports file system is full to the customer. [root@qehct1]# df -k Filesystem 1K-blocks Used Available Use% Mounted on /dev/sdb2 4182720 3982208 200512 96% / /dev/sdb1 101086 4798 91069 6% /boot none 1944112 0 1944112 0% /dev/shm /dev/sdb9 9765248 272 9764976 1% /usr/interchange /dev/sdb10 11717120 792 11716328 1% /data /dev/sdb11 37670048 9633924 28036124 26% /usr/g /dev/md0 143363584 100004448 43359136 70% /usr/g/sdc_image_pool
Error message
root: watchswap: Disk partition / is full (root partition) See long description cd /usr/adm tail –500 SYSLOG | more (this is the Octane computer system log) look for messages similar to this one identifying the root partition as being full Jan 14 12:12:24 3E:perryville_ct root: watchswap: Disk partition / is full Linux: cd /var/log more mess* Jul 9 04:35:56 ct01 logger: watchswap: Disk partition / is full (on ct01).
Applies to
- Models
- VCT
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 10.5_2.8.2I_H1.3M4 for LightSpeed 1.X (QXi) with Console Assembly (2180551) (Console with O2 Computer) or (2180551-2) (Console without O2 Computer) - showprods returns LightSpeed Apps ver 10.5_2.8.2I_H1.3M4 Released via FMI 25269 The features introduced with this software include(but are not limited to) the following:Thin Twin Axial Mode (2 x .63 mm) Multi-Destination DICOM Print Capability Latest release of Iniste Interactive (IIP 2.1) Advanced Artifact Reduction (AAR) Web Browser Based Common Service Desktop - Phase II Includes a new HomePage New Log Viewer Separated Class A, C, and M software Beam-Tracking Improvement LightSpeed Apps ver 10.5_2.8.2I_H1.3M4 Released via FMI 25269 The features introduced with this software include (but are not limited to) the following: Thin Twin Axial Mode (2 x .63 mm) Multi-Destination DICOM Print Capability Latest release of Iniste Interactive (IIP 2.1) Advanced Artifact Reduction (AAR) Web Browser Based Common Service Desktop - Phase II Includes a new HomePage New Log Viewer Separated Class A, C, and M software Beam-Tracking Improvement LightSpeed QX/i Rev 1.3 System Software Collector 2260334-5 consisting of 1.3 Software CD-ROM Collector 2258825-7 and Software Rating Plate 2240213-2 1.000 PC QX/I IRIX 65 VER 6.8.A DI 2240215-2 1.000 PC QX/I IRIX 65 VER 6.8.A DI 2240216-2 1.000 PC QX/I IRIX 65U VER 6U4.5.A 2258610-6 1.000 PC LIGHTSPEED QX/I APPS SFTW
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.